LuciPac A3 Filter Assay

ATP Test

How should I dispose of LuciPac?

No sterilization required. Dispose of according to local plastic disposal regulations. The main body is made of polypropylene, and glass and PVC (polyvinyl chloride) are not used. (For details, see the Lucipac Instruction Manual PDF [Disposal Method])"

Is it possible for the measured value to be "0"?

The tap water that wets the cotton balls also has ATP, ADP, and AMP, so the measured value never reaches "0". If the measured value becomes "0", there may be detergent, bleach, etc. that interfere with the measurement at the inspection site. It is also possible that there was a problem with the operation of the Lumitester.

How long is the expiration date of LuciPac?

Before opening, it is stable at 2-8℃ for 15 months after manufacturing date. The shelf life is printed on the bottom of the LuciPac aluminum bag.

After opening the aluminum bag, store it in a refrigerator and use it within 2 weeks.

Please also see the precautions for use.https://biochemifa.kikkoman.com/e/kit/method/atp-test-20/

The ATP Test (Kikkoman A3)

How should I dispose of LuciPac?

No sterilization required. Dispose of according to local plastic disposal regulations. The main body is made of polypropylene, and glass and PVC (polyvinyl chloride) are not used. (For details, see the Lucipac Instruction Manual PDF [Disposal Method])"

Is it possible for the measured value to be "0"?

The tap water that wets the cotton balls also has ATP, ADP, and AMP, so the measured value never reaches "0". If the measured value becomes "0", there may be detergent, bleach, etc. that interfere with the measurement at the inspection site. It is also possible that there was a problem with the operation of the Lumitester.

How long is the expiration date of LuciPac?

Before opening, it is stable at 2-8℃ for 15 months after manufacturing date. The shelf life is printed on the bottom of the LuciPac aluminum bag.

After opening the aluminum bag, store it in a refrigerator and use it within 2 weeks.

Please also see the precautions for use.https://biochemifa.kikkoman.com/e/kit/method/atp-test-20/

Does the Lumitester Smart need to be adjusted to zero point before each measurement?

No need to adjust the zero point for each measurement. AUTO ZERO calibration function is built-in for Lumitester Smart.

If the calibration for Lumitester is required by ISO or other regulations please contact your supplier you have purchased from. 

What can be the reason for the extremely low RLU (e.g., 0 RLU) results compared to usual. What is affecting the results?

There are 3 possible reasons that affects the measurements.

1) The test point or test sample contain alcohol, detergent, bleach, or other inhibitors, it may result lower RLU than usual. 

In some cases, it may show 0 RLU value if high concentrated detergent left on the test surface.

2) LuciPac was measured immediately after it was taken out of the refrigerator, it may also lower the RLU value.

3) Shutter malfunction in the Lumitester Smart. When there is a shutter malfunction inside the chamber, there is a possibility that it will effect the measurements.  

Why is ATP Test (Kikkoman A3) a effective hygiene monitoring system at food / beverage factory and health care center?

In food / beverage factory and health care center, it is important to quickly check the sanitary conditions.

The ATP Test (Kikkoman A3) measures the amount of ATP+ADP+AMP contained in soil/ residues and bacteria.

The cleanliness is instantly displayed as a numerical value of the tested area, on-site. 

What is the uniquness of ATP Test (Kikkoman A3) compared to the other ATP test?

ATP Test (Kikkoman A3) uses the unique enzymatic reactions in the proprietary A3 chemistry of our LuciPac reagents to change ADP and AMP to ATP enabling

the measurement of the total amount of ATP + ADP + AMP, producing a significantly more highly sensitive measurement for the detection of residual soil which can be missed when measuring ATP only.

Does detergent, bleach, sanitizer, disinfectant effect the RLU results?

Yes, it may. Any detergent, bleach (sodium hypochlorite), alcohol or other disinfectant, or disinfectant left on the surface may affect the results.

Degree of effect will depend on the inhibitory substances. If the inhibitor is rinsed thoroughly, there will be no effect in most cases.

For alcohol, there will be no effect when they are fully evaporated.

For more information: https://biochemifa.kikkoman.com/e/kit/method/atp-test-20/

*ATP Test (Kikkoman A3) is recommended to be performed after cleaning and before disinfection.

Can RLU counts be converted/calculaterd into CFU?

RLU reults cannot be replaced by the number of bacteria. The ATP Test (Kikkoman A3) measures the

ATP+ADP+AMP derived from soil/ residue and bacteria at the same time.

Even if the bacteria is present, the amount of ATP+ADP+AMP each bacteria has differs depending on the type and condition of the bacteria.

It is impossible to determine how many bacteria it corresponds. It may show corelation only when pippeting microorganisms and perform measurement in dilution series.

Is there correlation between the RLU and CFU?

There are no correlation between RLU and CFU plate counts for the typical condition/ surface.

There are microorganisms and ATP+ADP+AMP derived from soil/ residue when swabbing environmental surfaces. 

For more information: https://biochemifa.kikkoman.com/e/kit/method/atp-test-03/

Is there any soil / residue that ATP Test (Kikkoman A3) cannot measure?

Kikkoman A3 detects the amount of ATP+ADP+AMP. Any soil/ residue that doesn't include ATP, ADP and AMP would not be measured. 

*Viruses, which are not technically living cells, usually do not contain ATP

Kikkoman Biochemifa is closely monitoring an outbreak of coronavirus ( COVID-19 ).

*Apart from viruses, microorganisms and other organic substances contain ATP. Consequently, it is effective for daily evaluation of cleaning

Why does ATP need to be measured?